10.6084/m9.figshare.9632231.v2 Alok Prakash Alok Prakash Bindu Kumari Bindu Kumari Shiru Sharma Shiru Sharma A low-cost, wearable sEMG sensor for upper limb prosthetic application Taylor & Francis Group 2019 Surface electromyography myoelectric prosthesis sensitivity response time control 2019-09-11 15:40:47 Media https://tandf.figshare.com/articles/media/A_low-cost_wearable_sEMG_sensor_for_upper_limb_prosthetic_application/9632231 <p>This paper proposes a low-cost and sensitive surface electromyography (sEMG) sensor for the myoelectric prosthesis. The sensor consists of a skin interface, signal conditioning circuitry and power supply unit all encased in a single package. The tuned RC parameters based envelope detection scheme employed in the sensor enables faster as well as reliable recognition of EMG signal patterns regardless of its strength and subject variability. The output performance of the developed sensor was compared with a commercial EMG sensor regarding signal-to-noise ratio (SNR), amplitude sensitivity and response time. EMG signals with both the devices were acquired for 10 subjects (three amputees and seven healthy subjects), to perform this comparative analysis. The results showed 4× greater SNR values and 50% higher sensitivity of the developed sensor than the commercial EMG sensor. Also, the proposed sensor was 57% faster than the commercial sensor in producing the output response. The sensor was successfully tested on amputees for controlling a 3D printed hand prototype utilising a proportional control strategy. The enhanced output parameters of the sensor were responsible for smooth, faster and intuitive actuation of the prosthetic hand fingers.</p>