posted on 2019-09-16, 14:25 authored by Koen T van Hamersveld, Perla J Marang–van de Mheen, Lennard A Koster, Rob G H H Nelissen, Sören Toksvig-Larsen, Bart L Kaptein

Background and purpose — Pooling data of studies evaluating total knee arthroplasty migration using radiostereometric analysis (RSA) may be compromised when the RSA method used would influence estimated differences between groups. We therefore reanalyzed a marker-based RSA study with model-based RSA to assess possible limitations of each RSA method, including insert micromotions in modular TKA and their effect on estimated group differences.

Patients and methods — All patients had received a cemented Triathlon implant (Stryker, Mahwah, NJ, USA) with either an all-polyethylene (n = 29) or a metal-backed (n = 28) tibial component. The latter group was reanalyzed with model-based RSA. Precision of each RSA method was calculated using double examinations. Bland–Altman plots were constructed to determine the limits of agreement between the 2 RSA methods. Polyethylene insert micromotion was quantified by measuring migration with respect to the metal tray. Finally, analyses of the original study were repeated with the model-based RSA results.

Results — Systematic differences were found in translations between marker-based and model-based RSA as a result of different reference origins being used for migration calculations. Micromotions of the polyethylene insert within the metal tray were negligibly small. Mean migration results were comparable between marker-based and model-based RSA when using the same reference origin, even though conclusions on individual patients may differ between RSA methods due to various types of measurement error (e.g., marker occlusion and model-fit inaccuracies).

Interpretation — At least for the studied TKA design, pooling mean migration data of different RSA methods appears justified. For translations, however, adjustments should be made to correct for differences in reference origin. Migration patterns of individual patients may differ as a result of distinct types of measurement error.